Photorefraction (PR) methods have beenwidely used for pediatric vision screening since the 1980’s. While PR is easy to implement, the accuracy of refractive error measurements in humans has been unsatisfactory, largely due to the variations of intraocular scattering, the retinal reflectance, and pupil size. The objective of our studies was to clinically evaluate the accuracy of refraction measurements of an improved PR-based device, the Dynamic Ocular Evaluation System (DOES), and to investigate whether the accuracy is affected by the patient’s age, race, and pupil size, which are relevant to individual intraocular scattering and retinal reflectance. We performed DOES measurements in 99 volunteers (198 eyes) under two environmental light conditions and using four fixation targets. These results were compared to the standard clinical refraction testing performed the same day. The correlation and standard deviation were determined by Bland-Altman analysis. The influence of intraocular scattering, retinal reflectance, and pupil size was evaluated by comparing results from different age groups, races, and lighting conditions. In the region between -4 diopter (D) and +4D, the device showed a binocular refraction measurement accuracy of 0.45 D, 0.3 D, and 0.18 D root-mean-square (RMS) error (n=1337) for spherical equivalent and cross cylinders Jo and J45, respectively. Inaccuracy increased at high refraction (>4D). Age, race, and pupil size did not appear to significantly affect DOES PR measurement. This suggests that enhancements in the PR system and analysis may satisfactorily correct intersubject variability that currently limits the clinical use of PR devices and measurements.
Photoscreening (PS) has been a popular objective method for pediatric vision screening because of its simplicity of use
The name photorefraction (PR) refers to the refraction measurement made from PS images. PR was introduced in the early 1980s
PR analysis is based on the double-pass reflex image pattern that is reflected from retina. Factors that can contribute to the uncertainty of the analysis can include the individual’s retinal reflectance, the level of intraocular scattering, and the pupil size, which may vary with age and environment lighting conditions. In addition, many users point out that the PR systems were affected by the subject’s race
The PR optical theory is described previously
155 patients with an age range 4 to 81 years were recruited during their visits for comprehensive eye examinations for eye glasses or contact lens in Walmart Vision Center in Tullahoma, Tennessee. After obtaining informed consent, the volunteers were brought to a testing room to view through a binocular eye-piece a 3D cartoon that is located on a screen at 75 cm. Four refraction measurements were acquired using infrared DOES while the subjects viewed a fixation cartoon figure at 4 different stereo distances that ranged from 75-cm to infinity. To investigate the influences of environment lighting and the consequent pupil-size on the PR measurement, the test was performed twice under the conditions of normal room-light and lights-off. After the cartoon viewing, the patients were returned to their original eye exam schedule. Each eye would generate 8 sets of PR refraction data points under the two environment lighting and four visual stimuli conditions. For comparison, the clinical refraction data were collected. These include data from an autorefractor, which was an average result of three measurements, from the phoropter and trial lenses performed by the optometrist, and from cycloplegic retinoscopy for young children from whom subjective refraction were not feasible.
To prevent error from accommodation, eye data from 56 hyperopic or mildly myopic (Spherical Equivalent > -1 diopter) subjects who were younger than 50 years old were omitted in this refraction evaluation, leaving 99 subjects included in the refraction analysis.
In the PR measurements, 18.7% of tests were identified to contain images with poor quality due to eye movements or obstruction from the eyelids or eyelashes. Images were filtered with a quality control code based on the circularity, edge sharpness, and significant location-shifting. Since four detections were attempted in each measurement, the rejection rate of refraction measurements is 0.12% = (18.7%)4. In fact, all 198 eyes from the included 99 subject yielded qualified refraction results. In summary, the comparison was able to be made from the 99 subjects, 198 eyes, 1791 data points (OD= 888, OS=903; total n=1791) measured under 4 visual stimuli distances and 2 environment lighting conditions.
Age range: 7-81 years old.
Pupil range: 2.48 to 8.05 mm.
Refraction range: -13.5 to +6.1 diopter
We used both the correlation and the Bland-Altman statistical methods to evaluate agreement between DOES and the clinical gold-standard measurements that were obtained from the optometrist.
The correlation plots of spherical equivalent (SE) and crossed cylinders (J0 and J45) between the clinical gold standard (x-axis) and the DOES data (y-axis) are shown in the upper three panels of
The Bland-Altman analysis is shown in the three lower panels of
The monocular autorefractor is currently the most reliable objective refraction tool in the clinical setting. We also compared monocular autorefractor measurements to the clinical subjective refraction results in the same group of 99 patients (198 eyes). The standard errors of the SE were 0.36 diopter and crossed cylinders J0 and J45 were 0.14 and 0.18 diopters, respectively. The mean differences were also minimal (-0.09, 0.03, and -0.04, respectively). The 0.36 D error with 0.09 D mean error is a decent reference. It should be mentioned that each autorefractor data is an average of 3 consequent measurements, which automatically reduces the measurement inaccuracy in each individual. The DOES analysis in this section, however, includes every applicable measurement without taking the average.
The intraocular scattering normally increases with age and as cataracts develop. The onset age of age-related cataract is typically in the 40s. Intraocular scatteringis not only an important problem that reduces contrast of vision, it could also decrease the reliability of ophthalmic measurements. This is especially significant to PR since the measurement relies on the double pass light signal. To examine the influence of intraocular scattering on DOES accuracy, analysis was carried out in three age groups with similar distribution of refraction and pupil size. The details of these 3 age groups are:
1. 20 children from 7 to 15 years old with the total of 38 eyes and 377 tests. The distribution of refractive errors is SE= -1.8 ± 1.0 D. Pupil-size distribution is 5.8 ± 0.9 mm (range from 3.4 to 8.1 mm);
2. 24 young adults from 17 to 38 years old with total of 45 eyes and 434 tests. The distribution of refractive errors is SE= -1.9 ± 0.9 D. Pupil size distribution is 5.2 ± 1.3 mm (range from 2.9 to 7.2 mm);
3. 33 older adults from 41 to 81years old with total of 65 eyes, 525 tests. Distribution of refractive error is SE= +0.3 ± 1.8 D. Pupil size distribution is 4.7 ± 1.0 mm (range from 2.5 to 7.2 mm).
The result in SE, J0, J45, and the total focus errors from the standard exam were shown in
Pupil size variation has been one of the major difficulties in traditional PR or PS devices. We acquired DOES data in bright and dark environment settings to study the effect of pupil size in measurement accuracy. The older adults’ pupils did not undergo a large enough size change in response to room light conditions and could not be included in the analysis. Therefore, only children and young adult groups were considered in pupil size analyses. The subjects’ distributions in different pupil groups and the results are shown in
Total focus error was 0.47D in small pupils and 0.57D in large pupils in children. In adults, total focus error was 0.45D in small pupils and 0.40D in large pupils. There was no discernable effect of pupil size on accuracy in either age group. The difference of focus error was 0.1D or less between pupil sizes and the tendency was toward opposite directions. Thus, pupil size does not appear to be a significant factor in the DOES refraction measurement.
We compared measurement accuracy across races to investigate the influence of retinal reflectance on the double-pass PR technique. The group of 13 non-white subjects included 3 Asians, 3 Hispanics, and 7 blacks. The 42 white subjects were selected for the equivalent age, pupil size, and refraction distribution. The sample conditions of each group and the error analysis results are shown in
Pediatric refraction exam is very important, but is very restricted by the poor collaboration in this population. Photorefraction method provides an optional and objective approach for vision assessment. However, the PR measurement is not sufficiently precise and is generally considered as means to quickly estimate refraction. Enhancement of the technology and well understanding of the results could ease the diagnosis and treatment in clinical practice. This study describes the three major sources of the inaccuracy in PR measurement. Two modifications of a PR device, DOES, are made: a coaxial illumination to balance the individual’s retinal reflectance and a mathematic method to reduce the offset noise signal, which is assumed to be relevant to intraocular scattering. We evaluate the quantitative accuracy of DOES and examines the influence of age, race, and pupil size on measurement accuracy. We found that in the pediatric vision screening region between -4D and +4D, the device shows a binocular-refraction measurement accuracy that is indicated by standard errors 0.45D, 0.3 D, and 0.18 D (n=1337) for spherical equivalent and cross cylinders Jo and J45, respectively.
The accuracy of the modified PR device did not indicate significant influence by the race, age, pupil-size, or type of refraction. The measurement total focus error in the distinguished subject groups were fundamentally the same, around 0.48 +/- 0.06D. This suggests that the enhancements in the PR system and analysis may correct for intersubject variability that limits current PR devices and measurements.
The authors would like to acknowledge the UTRF maturation funding support, the support and consultation from Dr. Palmer and Walmart Vision Center at Tullahoma, Tennessee, and the support of Center for Laser Applications of the University of Tennessee Space Institute.